Preserving records when a breathalyzer's clock is wrong

6 min read

Safety managers and equipment staff who run breathalyzers may discover, while looking up stored test results, that the date and time on the device screen differ from the actual time. The problem here does not end with simply setting the clock. The time stored with a test result can be the basis for finding a particular result and for cross-checking it with other data the organization holds, such as duty rosters and vehicle trip records.

If you change the setting right after finding a time error, it becomes hard to know what time the device showed before the change and which period of records needs to be re-examined. To preserve records, you must separate correcting the current time from interpreting test results already stored. The key is not to rewrite past records arbitrarily, but to keep a continuous record of the state at the time of discovery and the actions taken afterward.

Record the screen and the reference used before making changes

When you find a time difference, first record the date and time shown on the device screen. Also note the actual date and time you checked the screen, who checked it, and the clock or system used for comparison. Rather than vaguely writing "the correct time," record specifically what the reference was, such as the organization's clock or an electronic clock you actually compared against, so the checking process can be reproduced later.

If possible, also keep material that shows the screen before the change. If there are stored test result files or printouts, leave them in their original state without deleting or overwriting them. This material does not only show that the device's clock was off. It becomes the basis for confirming what displayed time the results were stored under.

At this stage, it is best not to change the device time immediately. Once you change the setting, it is hard to check the earlier state again, and it can also become unclear which periods of past records need explanation. The order that best protects the records is to record the state first, then check the guidance for your device and decide whether to set the current time.

Do not mix observed facts with suspected causes

The fact that the device time differs from the actual time is a confirmed fact. Whether a recent battery replacement, repair, or long-term storage caused the time error, on the other hand, needs to be checked separately. If work was actually done recently, you can note the type of work and the date, but keep it as "recent work" rather than concluding it was the cause.

You also need to confirm whether the date on the device screen and the date shown in the stored files are on the same basis. Storage methods and time setting methods can differ by device, so from the screen display alone you cannot judge that the time in stored data was changed or offset in the same way.

As a hypothetical example, suppose that on the morning of June 10, you find that one device's screen time is behind the clock you compared it with. The observation is the device's screen time and the reference clock's time at the moment of discovery. The interpretation is that earlier test records may have the same difference, but when the difference began and whether it stayed constant have not been confirmed.

In this case, a possible action is to preserve the pre-change state and recent work history, and to look in the guidance for that model for how to set the time and any explanation about stored records. The completion standard is not simply that the current screen time is correct. It is a state in which you can distinguish before and after the setting change and have documented the range of past records that needs review. There is no need to conclude it was a battery or device fault while the cause remains unconfirmed.

Separate the original time from a separately interpreted time

You may need to tidy up times in internal reports or management logs. In that case, do not mix the original time stored by the device with a time a person wrote down separately after cross-checking other data in the same field. Keeping the original lets later staff tell which content is the device record and which is after-the-fact interpretation.

You should also be careful about applying a constant time difference to all past records at once. Without evidence that the difference was the same throughout that period, you cannot write the calculated time down as if it were the confirmed actual test time. Even when you cross-check with duty rosters or vehicle trip records, it is better to mark unconfirmed periods as they are and keep them linked to the original test results.

Category What to record Why to record it
State at the time of discovery Actual time checked and the device screen before the change Distinguishes when the time difference became known
Reference used for comparison The clock or system compared against and its time Shows on what basis the difference was judged
Setting change details Date and time of the change and who made it Separates records before and after the change
Past test data Files and printouts kept and the range reviewed Keeps original records and later interpretation from mixing

For example, if you compared a particular test result with trip records, write "device displayed time" and "time noted separately after cross-checking data" in separate fields in the management document. If you made a separate interpretation, also note the basis applied and the records it covers. This reduces the confusion of the next staff member rereading all past results based on the device's now-corrected time.

Decide on setting the current time from your model's guidance

How to set the time can differ by model. Applying another device's button sequence or settings screen directly can cause unintended changes, so first check that your device's model name matches the model covered by the guidance materials. DATECH's official Downloads center posts product materials such as the Korean-language manuals for The New AF-50 and the AF-50AD.

If you changed the current time following the documented procedure, compare the device screen against the reference time again right after the change and record the result. If the guidance does not make clear how stored results are affected, one option is to ask the manufacturer how to preserve the original data before changing the setting. Confirming how to set the time and confirming the basis for interpreting existing records are not the same task.

Confirming that the current screen time is correct is a confirmation of the state after setting. That alone does not mean the times of all past test records are accurate. Writing "current time corrected" and "range of past records reviewed" separately in the management document lets you communicate the scope of the correction without overstating it.

Keep repeated errors together with each device's history

If the time drifts again, do not record only that it was reset. Gather the screen time at each point of discovery, when the setting was changed, and any actual work done in between. Writing down verifiable history, such as battery replacement, repair, or storage, with dates becomes material for explaining the device's condition when inspection is needed later. This record is not for deciding the cause in advance, but for separating confirmed facts from unconfirmed matters.

If you run several breathalyzers together, divide the data by each device's serial number or internal management number. A time difference and setting change history found on one device cannot be applied to another. Only by keeping, for each device, the pre-change state, the change date, and the periods of past records that need explanation can you protect both the originality of test results and the subsequent management history.